型号
|
DOX 3104
|
DOX 3304
|
Screen
|
Colour 8’ TFT LCD screen, 800 x 480 pixels, 24 bits
Adjustment of brightness and contrast (500:1)
|
On-screen display
|
On 8x14 div with 4 channels + reference + Math functions and statistics table – full screen –
Vector or point modes with interpolation, permanent SPO mode: normal or colour
|
Language
|
French, English, Italian, Spanish and German - Help in French/English
|
Time base speed
|
100 MHz
|
300 MHz Bandwidth limit: 20 MHz
|
No. of channels
|
4 channels + 1 external channel
|
Max. input voltage
|
300 V (DC+AC Pk)
|
Vertical sensitivity
|
12 ranges from 2 mV to 10 V/div – Accuracy ±3% – 8-bit resolution
|
Rise time
|
< 3.5 ns
|
< 1.2 ns
|
Probe compensation factors
|
x 1 / 5 / 10 / 20 / 50 / 100 / 200 / 500 / 1,000
|
Time base speed
|
1 ns/div to 50s/div (oscilloscope)
|
Max. no. of traces captured per second
|
110,000 traces/s
|
Horizontal zoom
|
Compression, expansion
|
Automatic ROLL mode
|
From 100 ms/div to 50 s/div (1-2-5 step)
|
Sources/Mode
|
CH1, CH2 or CH3, CH4 Ext, Ext/5, AC line / Auto, Normal, One-shot
|
Type
|
Edge, Pulse (20 ns to 10 s), Amplitude (rise time, fall time), Video (NTSC, PAL, SECAM, HD and custom),
Windows, Interval, Dropout, Runt, Pattern
|
Trigger on serial bus and Decoding
|
I2C, SPI, UART
|
RS232, CAN, LIN
|
MSO logic analyser input
|
Option: 8 channels + clock for TTL/CMOS/LVCMOS3.3 and LVCMOS2.5/CUSTOM signals
|
Real-time sampling rate
|
ETS: 2GS/s
|
Vertical resolution
|
8 bits (vertical resolution 0.4%)
|
Acquisition depth
|
Up to 28 M: 14 Mpts per channel, adjustable: 7 k / 14 k / 70 k / 140 k / 700 k / 1.4 M / 7 Mpts
|
File manager
|
Trace files (DAV proprietary format and Excel-compatible “.CSV” format)
.SET configuration files – .BMP screenshot files
|
Acquisition
|
Normal, Peak detect, Average, High res
|
Display format
|
Y(t), Zoom, Roll, X-Y
|
“Statistics“ mode
|
Measurement of events
|
AUTOSET
|
AUTO adjustment: amplitude, time base and trigger
|
MATH function
|
Trace calculated in real time: CH1, CH2, CH3, CH4, + , - , x , / , (d/dt), integral (∫dt) and square root (√)
|
FF analyser
|
FFT calculated on 1,024 points - simultaneously with the waveform for the 4 channels
Adjustable windowing: rectangular, Hamming, Hanning, Blackmann
|
Cursors
|
Manual, Track mode and Auto
|
PASS/FAIL
|
Pass/Fail mode with specific terminal for envelope adjustment
|
Automatic measurements
|
32 measurements and statistics table
|
Built-in 25 MHz function generator
|
25 MHZ- 125 MS/s - 14 bits - arbitrary function generation with EasyWave on PC
|
Recording
|
Internal storage or USB flash drive on front panel
|
Printing
|
Via USB Device (PictBridge)
|
Communication on PC
|
Via USB device or Ethernet link with EASYSCOPE (OX) and EASYWAVE (GX) software
|
Power supply
|
Universal 100-240 V / 45-440 Hz/ 50 VAmax with removable cable
|
Safety / EMC / Locking
|
Compliant with the IEC 61010-1 standard, 300V CAT I - EMC as per EN61326-1 - Kensington lock
|
Temperature
|
Use: 0°C to +40°C, Storage:-20°C a +60°C
|
Mechanical specifications
|
352 x 111 x 224 mm – 3.6 kg (4 channels) – IP20 – 3-year warranty
|